Physics in Medicine and Biology · 1985 · 11 citations · 7 references
MedicineRadiologyRadiation DosimetryRadiation MeasurementRadiation ProtectionAbsorbed DoseExternal Beam DosimetryPhantom DosimetryQuality Assurance In DosimetryRadiation Dose MeasurementTreatment Planning Dose CalculationDose Distribution CalculationsRadiation Field CharacterizationRadiation Field Quality CorrectionWater Phantom Measurements
An analytical technique was employed to extract the primary component of dose at off-axis points from broad-beam water phantom data for a Clinac-18 Clinac-6 (100), Clinac-4 (Pb filter) and Philip's cobalt unit. The methodology presented here does not require separate off-axis HVL measurements and guarantees the accurate description of the data within the radiation field. It is shown that the off-axis reduction in the primary dose component for linacs does lead to significant (5-10%) errors in dose calculations if it is not taken into account in extended field external beam calculations. A method is described in which the off-axis variation in zero field TAR can be easily extracted from broad-beam scan data using the framework established by Cundiff et al (1973) as modified by Hanson and Berkeley (1980) for the dosimetry of large irregular fields. The technique is self-consistent and leads to parameters which have simple physical interpretations and which guarantee an accurate description of the beam.
7
A Note on Depth Doses in Fields of Irregular Shape
Jo Clarkson · British Journal of Radiology · 1941 · 227 citations
Medicine, Radiology, Medical Physics +14
J. R. Cunningham · Physics in Medicine and Biology · 1972 · 122 citations
A Note on the Calculation of Dose in X-ray Fields
Michelle Day · British Journal of Radiology · 1950 · 64 citations