Notation System and Overlay Diagrams for the Analysis of Paced Electrocardiograms

Alan D. Bernstein, Victor Parsonnet

Pacing and Clinical Electrophysiology · 1983 · 15 citations · 1 references

Concepts

Abstract

With the advent of adaptive dual-chamber pacing, the interaction between the pacemaker and the heart is becoming increasingly complex and the interpretation of paced electrocardiograms more difficult. We have developed a set of notation conventions for use in teaching and rhythm-strip analysis. Easily-drawn symbols delineate successful, unsuccessful, and inhibited pacing and sensing events as well as refractory and blanking periods. Another application is to generate movable transparent overlays after programming a pacemaker, to be used in interpreting subsequent rhythm strips.

References

1