The Analyst · 1996 · 26 citations · 34 references
Natural SciencesBiostatisticsStatistical ModelingStatisticsUncertainty QuantificationAnalytical CalibrationCalibrationConfidence IntervalsMeasurement ErrorMeasurement UncertaintyReliabilityCalibration StandardsError PredictionInstrument CalibrationInstrumental ErrorMeasurement Uncertainty EvaluationResponse ErrorSample InjectionStatistical Reliability Analysis
It is demonstrated that the statistical reliability of experimentally determined confidence intervals in some linear calibration problems in instrumental analyses can greatly be enhanced if the standard deviation (s, σM) of measurements, which is a theoretical prediction of the instrumental response error, is incorporated into the usual statistical equation instead of the residual of least-squares fitting. The reliability of the calibration depends on not only the fluctuation of calibration lines due to the response error, σM, but also the error of parametrization associated with the a priori response error prediction, i.e., the variance of the response s, Var(σM). The error prediction requires the Fourier transform of an instrumental baseline, signal shape and others (e.g., sample injection error). The variability in the confidence interval from five calibration standards is as small in the probabilistic approach as that for 50 standards in the statistical method. Liquid chromatography and capillary electrophoresis are taken as examples.
34
Principles of environmental analysis
Lawrence H. Keith, Warren B. Crummett, John Deegan et al. · Analytical Chemistry · 1983 · 755 citations
Decision and detection limits for calibration curves
Andre. Hubaux, G. Vos · Analytical Chemistry · 1970 · 514 citations