IEEE Sensors Journal · 2014 · 40 citations · 15 references
EngineeringBioengineeringBiomedical EngineeringBiomedical OpticsSensorsUrodynamicsFiber OpticsFiber-optic Medical DevicesOptical FibersPressure DistributionPressure MeasurementPressure SensorsAbdominal Pressure MeasurementBladder Pressure MeasurementFiber-optic Pressure SensorsFiber-optic ProbesFiber-optic SensorsInvasive UrodynamicsPressure SensitivityUrodynamic Pressure Measurement
A fiber optic sensing system for pressure measurement in invasive urodynamics is presented and experimentally validated. Probes are based on extrinsic Fabry-Perot interferometry (EFPI) principle, with an all-glass biocompatible design having 0.2-mm thickness. Pressure sensitivity of 1.0-1.6 nm/kPa with high stability and reduced thermal sensitivity has been achieved, leading to 0.3-cmH <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> O pressure accuracy. The EFPI probes have been embedded in medical catheters with 4F size for dual bladder and abdominal pressure measurement throughout the full course of a urodynamic analysis. Pressure traces have been compared with a PICO2000 urodynamic instrument, showing improved accuracy and higher responsivity to local pressure variations. Tests have been carried out in vivo on seven patients; the main highlights are reported, showing the premises for an EFPI-based diagnostic tool.
15
Good urodynamic practices: Uroflowmetry, filling cystometry, and pressure‐flow studies**
Werner Schäfer, Paul Abrams, Limin Liao et al. · Neurourology and Urodynamics · 2002 · 1.7K citations · Full text
Medicine, Urology, Female Urology +15