Concepedia

Publication | Closed Access

Fatigue to Fracture: An Informative, Fast, and Reliable Approach for Assessing Medical Implant Durability

24

Citations

6

References

2009

Year

Abstract

Abstract This paper compares the differences between the “test to success” and “test to fracture” paradigms for medical implant durability assessment. It recognizes that test to success can be one of the tests within test to fracture, in addition to a series of shorter cycle, higher force/deformation tests. In addition, by “testing to fracture,” one can validate the fatigue analysis capability, learn the consequences of in vitro fractures, and provide information that can either be used for product improvement or for clinicians to make life-saving decisions.

References

YearCitations

Page 1