Publication | Closed Access
Fatigue to Fracture: An Informative, Fast, and Reliable Approach for Assessing Medical Implant Durability
24
Citations
6
References
2009
Year
EngineeringFracture DiagnosticsMechanical EngineeringSurgeryBiomedical EngineeringOrthopaedic SurgeryFatigueFatigue ManagementShorter CycleOrthopaedicsDurability PerformanceHigher Force/deformation TestsImplantable DeviceFatigue Analysis CapabilityImplantologyLow-cycle FatigueDental BiomechanicsFracture HealingMedicineReliable Approach
Abstract This paper compares the differences between the “test to success” and “test to fracture” paradigms for medical implant durability assessment. It recognizes that test to success can be one of the tests within test to fracture, in addition to a series of shorter cycle, higher force/deformation tests. In addition, by “testing to fracture,” one can validate the fatigue analysis capability, learn the consequences of in vitro fractures, and provide information that can either be used for product improvement or for clinicians to make life-saving decisions.
| Year | Citations | |
|---|---|---|
Page 1
Page 1