Optics Letters · 1995 · 61 citations · 5 references
EngineeringBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMedical ImagingRadiographic ImagingX-ray DetectionX-ray DetectorsX-ray OpticsContrastContrast Sensitivity ImprovementLaser-produced Plasma X-raysMedical X-ray DetectionMicrochannel Plate DetectorsPlasma X-ray SourcesTemporal Contrast DiscriminationX-ray Imaging Physics
We demonstrate a contrast improvement of nearly a factor of 5 in a phantom medical x-ray image, using time-gated detection. This improvement is accomplished by temporal discrimination against scattered x-ray flux and may permit a significant reduction in patient x-ray exposure. A multiterawatt ultrashort-pulse laser-produced-plasma x-ray source (<1 ps) and a time-gated microchannel plate detector (~100 ps) are used.
5
Absorption of ultra-intense laser pulses
S. C. Wilks, W. L. Kruer, M. Tabak et al. · Physical Review Letters · 1992 · 1.8K citations
Natural Sciences, Physics, Optics +13
MeV x-ray generation with a femtosecond laser
J. D. Kmetec, C. L. Gordon, J. J. Macklin et al. · Physical Review Letters · 1992 · 374 citations
Natural Sciences, Physics, Optics +14
Multiterawatt 30-fs Ti:sapphire laser system
C. P. J. Barty, C. L. Gordon, B. E. Lemoff · Optics Letters · 1994 · 136 citations
Natural Sciences, Physics, Optics +14
Generation of x rays for medical imaging by high-power lasers: preliminary results.
K. Herrlin, G. Svahn, Carl Olsson et al. · Radiology · 1993 · 64 citations