EngineeringRadiologyInstrumentationRadiation DosimetryRadiation MeasurementDosimetry InstrumentationFilm DosimetryOptical InstrumentationRadiation Detector ArraysRadiation Imaging SystemsRadiation InstrumentationDetector Array SystemsDetector CalibrationFilm-based RadiographyImaging DetectorsOptical ScannersRadiographic DosimetryRadiographic Film DosimetrySpatial Response Surfaces
Film dosimetry is commonly performed by using linear CCD array transmission optical densitometers. However, these devices suffer from a variation in response along the detector array. If not properly corrected for, this nonuniformity may lead to significant overestimations of the measured dose as one approaches regions close to the edges of the scanning region. In this note, we present measurements of the spatial response of an AGFA Arcus II document scanner used for radiochromic film dosimetry. Results and methods presented in this work can be generalized to other CCD based transmission scanners used for film dosimetry employing either radiochromic or radiographic films.
9
Precise radiochromic film dosimetry using a flat-bed document scanner
Slobodan Dević, Jan Seuntjens, E Sham et al. · Medical Physics · 2005 · 569 citations