Temperature oscillation suppression of GM cryocooler

K Okidono, Takamasa Oota, Hideyuki Kurihara, Takeo Sumida, Takashi Nishioka, Hidemi Kato, M. Matsumura, Ouga SASAKI

Journal of Physics Conference Series · 2012 · 14 citations · 1 references

DOIFull text

Open access

Concepts

Abstract

GM cryocooler is a convenient refrigerator to achieve low temperatures about 4 K, while it is not suitable for precise measurements because of the large temperature oscillation of typically about 0.3 K. To resolve this problem, we have developed an adapter (He-pot) with a simple structure as possible. From the thermodynamic consideration, both heat capacity and thermal conductance should be large in order to reduce the temperature oscillation without compromising cooling power. Optimal structure of the He-pot is a copper cylinder filled with high pressure He-gas at room temperature. This can reduce the temperature oscillation to less than 10 mK below a certain temperature TH without compromising cooling power. TH are 3.8 and 4.5 for filled He-gas pressures of 90 and 60 atm, respectively. By using this He-pot, GM cryocooler can be applied to such as precise physical property measurements and THz detection.

References

1