Error in the calibration of the mgatp chemical‐shift limit: Effects on the determination of free magnesium by <sup>31</sup>P NMR spectroscopy

Timothy J. Mosher, Gerald D. Williams, Chris Doumen, Kathryn F. LaNoue, Michael B. Smith

Magnetic Resonance in Medicine · 1992 · 51 citations · 22 references

Concepts

Abstract

The measurement of free intracellular magnesium (Mg2+) using the 31P chemical shifts of ATP requires the use of appropriate calibration solutions to determine the chemical-shift limits delta ATP alpha beta and delta MgATP alpha beta. Solutions containing excess Mg2+ contain significant amounts of Mg2ATP and yield positive errors in the value of delta MgATP alpha beta. For physiological applications this may overestimate free intracellular Mg2+ by as much as 300%. This error may be minimized if appropriate mole ratios of Mg2+/ATP are used to calibrate delta MgATP alpha beta.

References

22