Measurement Science and Technology · 1996 · 16 citations · 7 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingDiagnostic Imaging SystemsMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingNeuroimaging PhysicsCoil DesignGenetic Algorithm Based DesignGenetic Algorithm OptimizationGradient Echo SequencesMagnetic Field Gradient SystemsPulse Sequence DesignStructural Brain Imaging
A genetic algorithm has been used to design a 35 cm diameter transverse gradient set for human brain imaging. Despite strict constraints on the maximum coil length (33 cm), gradient performance comparable to that from far longer sets of conventional geometry is achieved; thus the gradient efficiency is with a 24% standard deviation from this within a 26 cm sphere. Phase maps obtained using a gradient-acquisition for shimming sequence were used to test the linearity of the gradients without requiring a complex phantom.