American Journal of Roentgenology · 1986 · 39 citations · 0 references
Natural SciencesBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance Contrast MechanismsMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMedical ImagingNeuroimaging PhysicsCompressed Sensing Magnetic Resonance ImagingDynamic Magnetic Resonance Imaging ReconstructionDynamic Range CompressionField Strength DependenceWide Dynamic Range Compression
The large dynamic range in signal intensity present in MR surface-coil images makes proper windowing and photography difficult. By removing the low-spatial frequency information caused by variations in surface-coil field intensity from the high-spatial frequency information containing the image data, a considerable compression of this dynamic range of signal intensity is possible. To accomplish this, a technique was implemented on a digital computer for use with MR surface-coil image data. The compression was done as a postprocessing option after the patient scan had been completed and therefore did not alter actual scan times. Although the image signal-to-noise was not altered, the ease of photography for most images was improved. Thus, digital dynamic-range compression is a practical technique to aid in surface-coil MRI studies.