American Journal of Roentgenology · 1985 · 76 citations · 5 references
MedicineNeurologyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyIntervertebral Disc DiseaseMagnetic Resonance ImagingMedical ImagingMusculoskeletal ImagingNeurologic ImagingSpinal DisordersDegenerative Spine Disease ImagingHigh-resolution ImagingIntervertebral Disc ImagingLumbar Degenerative Disc DiseaseLumbar Spine DisordersLumbar Spine ImagingLumbar SpondylosisLumbosacral Disc Disease
Seventeen patients with lumbar disk disease were studied using a prototype magnetic resonance (MR) surface coil. The high signal-to-noise ratio achieved with the surface coil permitted increases in spatial resolution to 0.9 X 0.9 mm in-plane resolution with 5 mm slice thickness. The surface coil was also compatible with multiplanar, multiecho imaging techniques. The spatial resolution achieved in this study was nearly equivalent to that achieved by state-of-the-art computed tomographic (CT) scanners, and MR showed a superior range of soft-tissue contrast. One significant limitation of MR was its inability to demonstrate small calcifications. Nevertheless, MR imaging provided diagnostic information comparable to CT or myelography in a completely noninvasive manner. With further technical advances, MR is likely to become the initial procedure of choice for evaluating patients with suspected lumbar disk disease.
5
Nuclear magnetic resonance imaging of the spine.
Michael T. Modic, Meredith A. Weinstein, William Pavlicek et al. · Radiology · 1983 · 178 citations
Surface Coil Magnetic Resonance Imaging
Leon Axel · Journal of Computer Assisted Tomography · 1984 · 107 citations