Three‐dimensional flow‐independent peripheral angiography

Jean H. Brittain, Eric W. Olcott, Andrzej Szuba, Garry E. Gold, Graham A. Wright, Pablo Irarrázaval, Dwight G. Nishimura

Magnetic Resonance in Medicine · 1997 · 90 citations · 44 references

Concepts

Abstract

A magnetization-prepared sequence, T2-Prep-IR, exploits T1, T2, and chemical shift differences to suppress background tissues relative to arterial blood. The resulting flow-independent angiograms depict vessels with any orientation and flow velocity. No extrinsic contrast agent is required. Muscle is the dominant source of background signal in normal volunteers. However, long-T2 deep venous blood and nonvascular fluids such as edema also contribute background signal in some patients. Three sets of imaging parameters are described to address patient-specific contrast requirements. A rapid, spiral-based, three-dimensional readout is utilized to generate high-resolution angiograms of the lower extremities. Comparisons with x-ray angiography and two-dimensional time-of-flight angiography indicate that this flow-independent technique has unique capabilities to accurately depict stenoses and to visualize slow flow and in-plane vessels.

References

44