Review of Scientific Instruments · 2012 · 32 citations · 14 references
EngineeringNuclear EngineeringPhysicsElectromagnetismInstrumentationUncertainty QuantificationAccelerator PhysicsDetector TechnologyDigital InstrumentationSuperconducting Magnet SystemsAcceleratorsCurrent MeasurementField MeasurementsMagnet TestingMagnetic Field MeasurementsPrecision MeasurementSignal-to-noise RatioTime Resolution
The fast digital integrator has been conceived to face most demanding magnet test requirements with a resolution of 10 ppm, a signal-to-noise ratio of 105 dB at 20 kHz, a time resolution of 50 ns, an offset of 10 ppm, and on-line processing. In this paper, the on-field achievements of the fast digital integrator are assessed by a specific measurement campaign at the European Organization for Nuclear Research (CERN). At first, the architecture and the metrological specifications of the instrument are reported. Then, the recent on-field achievements of (i) ±10 ppm of uncertainty in the measurement of the main field for superconducting magnets characterization, (ii) ±0.02 % of field uncertainty in quality assessment of small-aperture permanent magnets, and (iii) ±0.15 % of drift, in an excitation current measurement of 600 s under cryogenic conditions, are presented and discussed.
14
Guide to the Expression of Uncertainty in Measurement
Barry N. Taylor, C. E. Kuyatt · 1995 · 821 citations
A. Jain · CERN Document Server (European Organization for Nuclear Research) · 1998 · 38 citations · Full text