Radiology · 1992 · 55 citations · 0 references
MedicineNeurologyRadiologyDiagnostic ImagingDiagnostic RadiologyNeuroradiologyBrain ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMedical ImagingNeurologic ImagingContrastFast Spin EchoMagnetic ResonanceMagnetic Resonance Imaging Lesion DetectionMagnetic Resonance PhysicsPulse SequencesSpin Echo ImagingSpin Echo Sequences
Fast spin-echo (FSE) magnetic resonance (MR) imaging was compared with conventional, peripherally gated T2-weighted spin-echo (SE) imaging in the detection of high- and low-signal-intensity lesions in the central nervous system. Lesion detectability was determined with percentage of contrast measurements and contrast-to-noise ratios with two different measurements for noise. All three measures of lesion detectability were similar. FSE and SE sequences were quantitatively equivalent in the detection of high-signal-intensity lesions. The SE sequence, however, was superior to the FSE sequence in the detection of small, low-signal-intensity lesions in the central nervous system caused by magnetic susceptibility effects.