Annals of Otology Rhinology & Laryngology · 1991 · 55 citations · 7 references
MedicineOtolaryngologyRadiologyCochlear ImplantsDiagnostic ImagingOtologyCochlear AnatomyCochlear DisordersCochlear ImplantationImplant TechnologyMagnetic Resonance ImagingSurgical ImplantationImaging For Cochlear ImplantationMagnetic Resonance
Magnetic resonance imaging poses little risk to the majority of patients undergoing scanning. However, cochlear implantation should pose a contraindication to this imaging process. This conclusion is based on several findings of in vitro testing of three cochlear implants: the 3M/House and 3M/Vienna designs and the Nucleus device. Specifically, tremendous torques are generated by each of these devices when they are introduced into the coil of a magnetic resonance imager; in addition, the 3M products not only were noted to induce an electrical current, but also were significantly magnetized and rendered afunctional. The clinical implications are discussed.
7
The effects of nuclear magnetic resonance on patients with cardiac pacemakers.
William Pavlicek, Michael Geisinger, Lon Castle et al. · Radiology · 1983 · 208 citations
Medicine, Cardiology, Radiology +17