Electrocautery can be used safely to create a tunical incision with any underlying inflatable cylinder. To avoid electrocautery injury, based on the clinical study results in polyurethane elastomer lined devices, one should deflate the cylinder before electrocautery, use coagulation current at 35 watts, apply the electrocautery only to the outer longitudinal tunical layer, bluntly dissect through the inner circular layer, and elevate the tunica, protect the device and incise the tissue under direct vision.