A Clinical and Experimental Comparison of Time of Flight PET/MRI and PET/CT Systems

Daniela E. Oprea‐Lager, Maqsood Yaqub, Indra Pieters, Rinze Reinhard, R. Jeroen A. van Moorselaar, Alfons J.M. van den Eertwegh, Otto S. Hoekstra, Adriaan A. Lammertsma, Ronald Boellaard

Molecular Imaging and Biology · 2015 · 14 citations · 34 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: The purpose of the study was to compare image quality and quantitative accuracy of positron emission tomography/magnetic resonance imaging (PET/MRI) and PET/computed tomography (PET/CT) systems with time of flight PET gantries, using phantom and clinical studies. PROCEDURES: Identical phantom experiments were performed on both systems. Calibration, uniformity, and standardized uptake value (SUV) recovery were measured. A clinical PET/CT versus PET/MRI comparison was performed using [(18)F]fluoromethylcholine ([(18)F]FCH). RESULTS: Calibration accuracy and image uniformity were comparable between systems. SUV recovery met EANM/EARL requirements on both scanners. Thirty-four lesions with comparable PET image quality were identified. Lesional SUVmax differences of 4 ± 26% between PET/MRI and PET/CT data were observed (R (2) = 0.79, slope = 1.02). In healthy tissues, PET/MRI-derived SUVs were 16 ± 11% lower than on PET/CT (R (2) = 0.98, slope = 0.86). CONCLUSION: PET/MRI and PET/CT showed comparable performance with respect to calibration accuracy, image uniformity, and SUV recovery. [(18)F]FCH uptake values for both healthy tissues and lesions corresponded reasonably well between MR- and CT-based systems, but only in regions free of MR-based attenuation artifacts.