Publication | Closed Access
Effect of tube current modulation for dose estimation using a simulation tool on body CT examination
13
Citations
13
References
2014
Year
Computed TomographyEngineeringTube Current ValuesSurgeryBody Ct ExaminationDual-source CtTreatment VerificationCt ScanRadiation Therapy PlanningPhoton-counting Computed TomographyNuclear MedicineRadiologyTube Current ModulationMedical ImagingUltrasoundRadiographic ImagingDosimetryBiomedical ImagingDose EstimationMedicineAnesthesiology
The purpose of this study was to evaluate the effect of tube current modulation for dose estimation of a body computed tomography (CT) examination using a simulation tool. The authors also compared longitudinal variations in tube current values between iterative reconstruction (IR) and filtered back-projection (FBP) reconstruction algorithms. One hundred patients underwent body CT examinations. The tube current values around 10 organ regions were recorded longitudinally from tube current information. The organ and effective doses were simulated by average tube current values and longitudinal modulated tube current values. The organ doses for the bladder and breast estimated by longitudinal modulated tube current values were 20 % higher and 25 % lower than those estimated using the average tube current values, respectively. The differences in effective doses were small (mean, 0.7 mSv). The longitudinal variations in tube current values were almost the same for the IR and FBP algorithms.
| Year | Citations | |
|---|---|---|
Page 1
Page 1