Left ventricular volume from paired biplane two-dimensional echocardiography.

N B Schiller, Harry Acquatella, Thomas A. Ports, DENIS DREW, J Goerke, Hans Ringertz, Norman H. Silverman, Bruce H. Brundage, E H Botvinick, Rosaleen Boswell,

Circulation · 1979 · 610 citations · 0 references

Concepts

Abstract

To evaluate the applicability of two-dimensional echocardiography to left ventricular volume determination, 30 consecutive patients undergoing biplane left ventricular cineangiography were studied with a wide-angle (84 degrees), phased-array, two-dimensional echocardiographic system. Two echographic projections were used to obtain paired, biplane, tomographic images of the left ventricle. We used the short-axis view (from the precordial window) as an anolog of the left anterior oblique angiogram, and the long-axis, two-chamber view (from the apex impulse window) as a right anterior oblique angiographic equivalent. A modified Simpson's rule formula was used to calculate systolic and diastolic left ventricular volumes from the biplane echogram and the biplane angiogram. These methods correlated well for ejection fraction (r = 0.87) and systolic volume (r = 0.90), but only modestly for diastolic volume (r = 0.80). These correlations are noteworthy because 65% of the patients had significant segmental wall motion abnormalities. The volumes determined from the minor-axis dimensions of M-mode echograms in 23 of the same patients correlated poorly with angiography.