A quantitative evaluation of various iterative deconvolution algorithms

P.B. Crilly

IEEE Transactions on Instrumentation and Measurement · 1991 · 103 citations · 10 references

Concepts

Abstract

Various iterative deconvolution algorithms are evaluated that are commonly used to restore degraded chromatographic or spectroscopic peak data. The evaluation criteria include RMS errors, relative errors in peak areas, peak area variances, and rate of convergence. The iterative algorithms to be evaluated include Van Cittert's method, Van Cittert's method with constraint operators, relaxation based methods, and Gold's ratio method. The discussion also includes some enhancements that will improve the algorithms' convergence properties.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

10