Accidental cross‐connexion of oxygen and nitrous oxide in an anaesthetic machine

Augustine K. Bonsu, ALAN L. STEAD

Anaesthesia · 1983 · 18 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

Oxygen and nitrous oxide pipes were cross-connected during the servicing of an anaesthetic machine. This resulted in the delivery of 100% nitrous oxide instead of 100% oxygen at the end of the operation. The patient fully recovered and was discharged on the 13th day.