MTF, NPS and DQE characterization of an in-house developed X-ray imaging detector for synchrotron based micro-tomography

Kewin Desjardins, Michel Bordessoule, C. M. Petrache, Claude Menneglier, D. Dallé, Pascal Mercère, Kadda Medjoubi

Journal of Instrumentation · 2014 · 10 citations · 7 references

Concepts

Abstract

Hard X-ray micro-tomography is a powerful tool to reveal the internal structure of thick objects in a non-destructive manner. For synchrotron applications and practical lab works with students, a 2D detector based on a thin YAG:Ce scintillator optically coupled to a new generation high frame rate (100 fps) low noise sCMOS camera has been developed and characterized in depth on the METROLOGIE and PSICH'E beamlines of the SOLEIL synchrotron. The Detector gain, Modulation Transfer Function, Noise Power Spectrum and Detective Quantum Efficiency have been measured and compared with analytical model. A tomography reconstruction was performed on small insects to demonstrate the performance of this X-ray imaging detector.

References

7