Annals of Clinical Biochemistry International Journal of Laboratory Medicine · 2010 · 27 citations · 7 references
Interference in immunoassay caused by endogenous immunoglobulin is a cause of incorrect laboratory results that can drastically affect patient management. Two cases of immunoglobulin interference in serum follicle-stimulating hormone (FSH) assays are presented. These cases illustrate two common mechanisms for false-positive interference in two-site (sandwich) immunoassays. The first case describes a circulating autoimmune FSH immunoglobulin complex ('macro'-FSH), which has not been previously described for FSH, and the second a cross-linking antibody directed against the assay reagents. Immunoglobulin interference was detected and characterized using a combination of method comparison, immunosubtraction and size exclusion chromatography.
7
Michael Fahie‐Wilson, Sara Soule · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · 1997 · 185 citations
Immunology, Diagnosis, Pathology +17
Specificity and Clinical Utility of Methods for the Detection of Macroprolactin
Lucille Kavanagh, T. J. McKenna, Michael Fahie‐Wilson et al. · Clinical Chemistry · 2006 · 112 citations · Full text