Reader and platform reproducibility for quantitative assessment of carotid atherosclerotic plaque using 1.5T Siemens, Philips, and General Electric scanners
47
Citations
17
References
2007
Year
All three scanners produced images that allowed AHA-LT to be consistently identified. Reproducibility of quantitative measurements by Siemens and Philips scanners were comparable to previous studies using 1.5T GE scanners. However, bias was introduced with each scanner and the use of different readers substantially increased variability. We therefore recommend using the same platform and the same reader for scans of individual subjects undergoing serial assessment of carotid atherosclerosis.