Procedia Engineering · 2015 · 42 citations · 3 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBiomechanicsFinite Element AnalysisOrthopedic BiomechanicsBone MechanicsExternal Skeletal FixationFixation BiomechanicsFracture BiomechanicsFracture FixationFracture MechanicsAxial Compression ResponseBiomechanics Of External FixationExternal Fixation MechanicsFixation Construct BiomechanicsFixation StiffnessFracture HealingStiffnessUnilateral External Fixators
This paper presents research results of a stiffness analysis of a Sarafix external fixation system. The stiffness property of the external fixator affects the local biomechanical and biological environment of fracture healing. The research has been conducted in a case when one of the Sarafix unilateral biplanar fixator configurations has been applied to a tibia with an open fracture. A stiffness analysis was performed using FEA and experimental testing under three types of loads: axial compression, AP (anterior-posterior) four-point bending and torsion. 3D geometrical and FEM model of the fixator configuration was formed using CATIA V5 software system and a structural analysis was performed afterwards. Verification of the results obtained from a structural analysis was carried out through experimental testing by comparing values of an appropriate component of displacement at the point of load.
3