Publication | Closed Access
Histopathologic Findings in Early Encapsulated Blebs of Young Patients Treated With the Ahmed Glaucoma Valve
61
Citations
0
References
2010
Year
Encapsulation of the AGV is an early complication in young patients, leading to inhibition of fluid exchange and failure of the procedure. The valve mechanism is blocked by contracted scar tissue, but the device itself is not affected by the encapsulation. Surgical excision of the capsule immediately leads to an aqueous flow and drop of intraocular pressure.