Evaluation of RF Heating on Humerus Implant in Phantoms during 1.5T MR Imaging and Comparisons with Electromagnetic Simulation

25

Citations

17

References

2006

Year

Abstract

RF heating was verified even in a nonmagnetizing metal implant in a case of excessive RF irradiation. Particularly, rapid temperature rise was observed at both ends of the implant having large curvatures. The difference in temperature increase by depth was found to reflect the skin-depth effect of RF intensity. Electromagnetic simulation was extremely useful for visualizing the eddy currents within the loop and loop-cut phantoms and for evaluating RF heating of a metallic implant for MRI safety.