Ultrasonic Power/Data Telemetry and Neural Stimulator With OOK-PM Signaling

Ye-Sing Luo, Jiun-Ru Wang, Wei‐Jen Huang, Je-Yu Tsai, Yi-Fang Liao, Wan-Ting Tseng, Chen‐Tung Yen, Pai‐Chi Li, Shen-Iuan Liu

IEEE Transactions on Circuits & Systems II Express Briefs · 2013 · 28 citations · 18 references

Concepts

Abstract

An ultrasonic power/data telemetry and a neural stimulator are presented. An on-off keying with pulse modulation is adopted for ultrasonic signal transmission. The ultrasonic power is used to charge a battery. The ultrasonic data are used for neural stimulation, and an animal study is also demonstrated. The range of the stimulation current is 0-640 μA. The frequency of the stimulation current pulse is from 60 to 265 Hz. This ultrasonic power/data telemetry with a neural stimulator is fabricated in a 0.35- μm CMOS technology. It occupies the area of 4.4 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The total power dissipation is 2 mW with a stimulation current of 140 μA and a stimulation frequency of 265 Hz.