Applied Physics Letters · 2006 · 178 citations · 8 references
EngineeringChemistryAnalytical ChemistryNuclear InstrumentationSensorsMagnetic CharacterizationMagnetic SensorsExplosivesExplosives DetectionField SensitivityMagnetometer DetectionMagnetometer MeasurementsNuclear Magnetic Resonance SpectroscopyNuclear Quadrupole ResonanceSignal-to-noise Ratio
A radio-frequency tunable atomic magnetometer is developed for detection of nuclear quadrupole resonance (NQR) from room temperature solids. It has a field sensitivity 0.24fT∕Hz1∕2 at the 423kHz N14 NQR frequency of ammonium nitrate. A potential application of the magnetometer is detection of nitrogen-containing explosives which is difficult with conventional tuned copper coils due to a poor signal-to-noise ratio (SNR) below a few megahertz. The NQR signal from 22g of powdered ammonium nitrate located 2cm away from the sensor is detected with a SNR of 9 in a 4.4-s-long multiple echo sequence, which represents an estimated order-of-magnitude improvement in sensitivity over the pickup coil detection.
8
NMR Detection with an Atomic Magnetometer
Igor Savukov, M. V. Romalis · Physical Review Letters · 2005 · 248 citations · Full text
Free Magnetic Induction in Nuclear Quadrupole Resonance
M. Bloom, E. L. Hahn, B. Herzog · Physical Review · 1955 · 183 citations