A Selenium-Based Detector System For Digital Slot-Radiography

Walter Hillen, U. Schiebel, T. Zaengel

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 1988 · 13 citations · 0 references

Concepts

Abstract

A research system for digital radiography is described, which is based on a selenium detector with capacitive probe readout. The detector, in which a selenium drum is used as the primary image receptor, is exposed by a scanning fan beam. Scatter reduction and primary transmission by slot-radiography as well as the imaging properties of the selenium detector are discussed. The spatial resolution and the noise behaviour of the detector are analysed. The signal-to-noise ratios expressed in terms of noise equivalent quanta and detective quantum efficiency are calculated and compared with competitive systems.