Review of Scientific Instruments · 1979 · 31 citations · 6 references
Natural SciencesElectrical EngineeringPhysicsInstrumentationNuclear InstrumentationPlasma PhysicsDetector TechnologyHigh-temperature PlasmasPlasma DiagnosticsPlasma ElectrodynamicsCyclotron RadiationElectron Cyclotron Emission DiagnosticsHeterodyne DetectionHeterodyne ReceiversPlasma Source DiagnosticsTemperature Profile Analysis
We have developed a fast-scanning heterodyne receiver into a plasma diagnostic that measures the fundamental cyclotron emission from the PLT plasma and thus ascertains the time evolution of the electron temperature profile. The receiver scans 60-90 GHz every 10 ms and is interfaced to a computer for completely automated calibrated temperature measurements.
6
Electron Cyclotron Emission from a Tokamak Plasma: Experiment and Theory
A. E. Costley, R. J. Hastie, J. W. M. Paul et al. · Physical Review Letters · 1974 · 185 citations
Electron cyclotron emission in Alcator tokamak
I. H. Hutchinson, D.S. Komm · Nuclear Fusion · 1977 · 91 citations
Turbulent Temperature Fluctuations in the Princeton Large Tokamak Plasma
V. Arunasalam, R. Cano, J. C. Hosea et al. · Physical Review Letters · 1977 · 30 citations