Journal of Feline Medicine and Surgery · 2013 · 56 citations · 29 references
Veterinary MedicineVeterinary PhysiologyFeline Internal MedicinePhysiatryUltrasonographyBlood Pressure ControlCardiovascular AssessmentCardiovascular MonitoringHemodynamicsArterial Blood Pressure MeasurementBlood PressureBlood Pressure MeasurementBlood Pressure MonitoringMeasurementNoninvasive Blood Pressure MeasurementRadio TelemetryValidation Criteria Development
This study compared indirect blood pressure measurements using a non-invasive method, high-definition oscillometry (HDO), with direct measurements using a radio-telemetry device in awake cats. Paired measurements partitioned to five sub-ranges were collected in six cats using both methods. The results were analysed for assessment of correlation and agreement between the two methods, taking into account all pressure ranges, and with data separated in three sub-groups, low, normal and high ranges of systolic (SBP) and diastolic (DBP) blood pressure. SBP data displayed a mean correlation coefficient of 0.92 ± 0.02 that was reduced for low SBP. The agreement level evaluated from the whole data set was high and slightly reduced for low SBP values. The mean correlation coefficient of DBP was lower than for SBP (ie, 0.81 ± 0.02). The bias for DBP between the two methods was 22.3 ± 1.6 mmHg, suggesting that HDO produced lower values than telemetry. These results suggest that HDO met the validation criteria defined by the American College of Veterinary Internal Medicine consensus panel and provided a faithful measurement of SBP in conscious cats. For DBP, results suggest that HDO tended to underestimate DBP. This finding is clearly inconsistent with the good agreement reported in dogs, but is similar to outcomes achieved in marmosets and cynomolgus monkeys, suggesting that this is not related to HDO but is species related. The data support that the HDO is the first and only validated non-invasive blood pressure device and, as such, it is the only non-invasive reference technique that should be used in future validation studies.
29
Recommendations for Blood Pressure Measurement in Humans and Experimental Animals
Thomas G. Pickering, John E. Hall, Lawrence J. Appel et al. · Hypertension · 2004 · 2.1K citations · Full text
Scott A. Brown, Clarke E. Atkins, R. Bagley et al. · Journal of Veterinary Internal Medicine · 2007 · 644 citations · Full text
Veterinary Medicine, Veterinary Clinical Sciences, Veterinary Internal Medicine +7
Scott A. Brown, Clarke E. Atkins, R. Bagley et al. · Journal of Veterinary Internal Medicine · 2007 · 559 citations · Full text
Veterinary Medicine, Veterinary Clinical Sciences, Veterinary Internal Medicine +7
Recommendations for Blood Pressure Measurement in Humans and Experimental Animals
Theodore W. Kurtz, Karen A. Griffin, A. Bidani et al. · Hypertension · 2004 · 262 citations · Full text