EngineeringMeasurementFetal MedicineWearable TechnologyAutocorrelation EstimationFetal ParametersPatient MonitoringBiostatisticsPublic HealthCardiologyCardiovascular ImagingBetter UseMaternal HealthPrenatal DiagnosisMaternal-fetal MedicineUltrasoundRadarCardiovascular DiseaseNew EstimatorsDoppler Ultrasound DevicePediatricsDoppler Ultrasound DevicesHealth MonitoringIntrapartum Ultrasound
Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5%, corresponding to an improvement of 14% compared to the standard estimator.