Evaluation of the EyeSys Model II Computerized Videokeratoscope. Part I: Clinical Assessment
19
Citations
0
References
1998
Year
EyeSys simulated keratometry values were not interchangeable with the keratometer. This could be due to differences in the method of alignment or measurements from different areas on the corneal surface. Repeatability of both the Bausch & Lomb keratometer and the EyeSys CAS was found to be similar. The spatial dependency of peripheral corneal radius repeatability measurements could be attributed to interference of the ocular adnexa in those areas of measurement.