Concepedia

Publication | Open Access

The effect of voxel size on high‐resolution peripheral computed tomography measurements of trabecular and cortical bone microstructure

114

Citations

30

References

2012

Year

Abstract

This study demonstrates the effect of voxel size on the accuracy of HR-pQCT measurements of trabecular and cortical microstructure and presents parameters for HR-pQCT analysis at nonstandard resolutions. For all parameters measured, correlations were strongest at 41 μm. Weak correlations for porosity measures indicate that a better understanding of pore structure and resolution dependence is needed.

References

YearCitations