Low outgassing residual gas analyzer with a beryllium–copper-alloy-flanged ion source

Fumio Watanabe, Akinari Kasai

Journal of Vacuum Science & Technology A Vacuum Surfaces and Films · 1995 · 18 citations · 0 references

Concepts

Abstract

By using a newly developed beryllium–copper (BeCu)-alloy ConFlat flange to house the hot-cathode ion source, a remarkable decrease in the outgassing from a quadrupole residual gas analyzer (RGA) has been achieved. The reduction in outgassing between the new BeCu-flanged RGA and an ordinary stainless-steel RGA of otherwise similar design was a factor of 60 or more in the 10−9 Pa total pressure range. From these results, the possibility of high accuracy residual gas analysis below 10−9 Pa is introduced.