Journal of Vacuum Science & Technology A Vacuum Surfaces and Films · 1995 · 18 citations · 0 references
EngineeringNuclear EngineeringAnalytical ChemistryInstrumentationNuclear InstrumentationDetector SystemsIonization TechniquesFlange TechnologyGas Analyzer DesignGas DetectionGas Detection InstrumentsGas Detection TechnologyIon Source PhysicsTotal Pressure
By using a newly developed beryllium–copper (BeCu)-alloy ConFlat flange to house the hot-cathode ion source, a remarkable decrease in the outgassing from a quadrupole residual gas analyzer (RGA) has been achieved. The reduction in outgassing between the new BeCu-flanged RGA and an ordinary stainless-steel RGA of otherwise similar design was a factor of 60 or more in the 10−9 Pa total pressure range. From these results, the possibility of high accuracy residual gas analysis below 10−9 Pa is introduced.