Publication | Closed Access
Personal monitoring of<sup>218</sup>Po and<sup>214</sup>Po radionuclide deposition onto individuals under normal environmental exposure conditions
14
Citations
12
References
1999
Year
Environmental MonitoringEngineeringRadioactive ContaminationExposure AssessmentDetector SurfaceEnvironmental HealthTastrak Alpha-sensitive PlasticToxicologyAnalytical ChemistryInstrumentationRadiation OncologyNuclear MedicineRadiologyHealth SciencesRadiation DetectionHuman ExposurePersonal MonitoringCosmic RayEcotoxicologyRadiation EffectsDosimetryEnvironmental EngineeringEnvironmental ToxicologyPersonal Dosemeters
Personal dosemeters have been utilized to monitor the deposition of the radon decay products 218Po and 214Po onto individuals under normal environmental exposure conditions. Each detector consists of TASTRAK alpha-sensitive plastic incorporated into an ordinary working wristwatch. Subsequent analysis provides energy discrimination of the detected alpha-particle decays, and allows events from the individual radon decay products 218Po and 214Po, attached to the detector surface, to be uniquely identified. Assuming similar deposition onto skin and detector surfaces, the activity per unit area of deposited radionuclides can be determined for exposed skin.