IEEE Sensors Journal · 2011 · 84 citations · 11 references
EngineeringBioengineeringBiomedical EngineeringBiomedical Signal ProcessingBiosensorsMicrowave EngineeringMedical InstrumentationMicrowave MeasurementsMicrowave Passive ComponentsVital SignsVital Signs AssessmentWearable Health TechnologiesExperimental Study DesignLeakage PowerMicrowave Phase Shifter DesignMillimeter-wave Integrated CircuitsWaveguide ComponentsWearable Health Monitoring
A first reported experimental study of a 60 GHz millimeter-wave life detection system (MLDS) for noncontact human vital-signal monitoring is presented. This detection system is constructed by using V-band millimeter-wave waveguide components. A clutter canceller is implemented in the system with an adjustable attenuator and phase shifter. It performs clutter cancellation for the transmitting power leakage from the circulator and background reflection to enhance the detecting sensitivity of weak vital signals. The noncontact vital signal measurements have been conducted on a human subject in four different physical orientations from distances of 1 and 2 m. The time-domain and spectrum waveforms of the measured breathing and heartbeat are presented. This prototype system will be useful for the development of the 60-GHz CMOS MLDS detector chip design.
11
Time-harmonic electromagnetic fields
C.W. Hargens · Journal of the Franklin Institute · 1961 · 2.9K citations
An X-Band Microwave Life-Detection System
Kun‐Mu Chen, Devendra K. Misra, Huei Wang et al. · IEEE Transactions on Biomedical Engineering · 1986 · 282 citations
Engineering, Biomedical Engineering, Electrical Engineering +17