Development of a Traceable Calibration Methodology for Solid<sup>68</sup>Ge/<sup>68</sup>Ga Sources Used as a Calibration Surrogate for<sup>18</sup>F in Radionuclide Activity Calibrators

Brian E. Zimmerman, Jeffrey T. Cessna

Journal of Nuclear Medicine · 2010 · 64 citations · 9 references

DOIFull text

Open access

Abstract

The ability for NIST to calibrate these epoxy-based mock syringes enabled, for the first time to our knowledge, the direct traceability to the national (68)Ge standard to be established for this type of source. Through a direct comparison with the NIST (18)F standard, the determination of the relative response ratios in activity calibrators enabled the equivalent (18)F activity to be determined in a way that was also traceable to the national (18)F activity standard.

References

9