IEEE Transactions on Biomedical Engineering · 1990 · 36 citations · 12 references
MedicineBiomedical EngineeringCardiologyDiagnostic ImagingUltrasonographyUltrasoundCardiac ImagingCardiac UltrasonographyCardiac UltrasoundTransesophageal EchocardiographyUltrasonicsUltrasound InstrumentationAccuracyLeft Ventricular ImagingOffline Data ProcessingQuantitative Volume MeasurementUltrasonic ImagingUltrasound Image AcquisitionVolumetric Data AcquisitionVolumetric Ultrasound Acquisition
We have developed a system to assess the feasibility of using multiple transesophageal ultrasonic images to measure left-ventricular volume, an important variable in patient management. The system includes a special transesophageal probe with a micromanipulator for acquiring cardiac images in multiple planes with known interplanar spatial relationship and an off-line processing system to compute the volume. In vitro studies with the probe demonstrated that the distance between two targets in space can be identified within 2 mm (SD = 0.4 mm) for points in the imaging plane 3.4 mm (SD = 0.5 mm) for points not lying in the imaging plane. This gives an average accuracy of +/- 6.5% for distances greater than 4.5 cm. Comparison of ultrasonic measurements of the volume of water-filled balloons and excised hearts to the volume required to fill them, revealed a correlation coefficient of 0.992, a regression line having a slope of 1.0 and an ordinate intercept at 0.2 mL, and a standard error of the estimate of 8 mL.