Rupture of the NovaFlex balloon during transcatheter aortic valve implantation and subsequent dissection of the right iliac arteries

Mehmet Gül, Hale Ünal Aksu, Kürşat Öz

European Journal of Cardio-Thoracic Surgery · 2012 · 13 citations · 3 references

DOIFull text

Open access

Concepts

Abstract

We report a case with a very rare complication of transcatheter aortic valve implantation. Rupture of the NovaFlex balloon (Edwards transfemoral balloon catheter) occurred during the inflation of the Edwards SAPIEN valve, resulting in dissection of the right common and external iliac arteries during withdrawal of the balloon catheter. The NovaFlex balloon is a part of the Edwards NovaFlex XT transfemoral delivery system.

References