Biomedical Chromatography · 2010 · 18 citations · 22 references
PharmacyPharmacokineticsPharmacologyAnalytical ChemistryBioavailabilityClinical PharmacologyAnalytical ChromatographyBioanalytical Method ValidationChromatographyInstrumental AnalysisLiquid ChromatographyMass SpectrometryAnalytical Liquid ChromatographyHigh-performance Liquid Chromatography Method DevelopmentHigh-performance Liquid Chromatography-mass SpectrometryLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLiquid Chromatography-mass Spectrometry Method DevelopmentRelative Oral BioavailabilityTandem Mass Spectrometry Analysis
The validation of a high throughput and specific method using a high-performance liquid chromatography coupled to electrospray (ES+) ionization tandem triple quadrupole mass spectrometric (LC-ESI-MS/MS) method for ondansetron quantification in human plasma is described. Human plasma samples were extracted by liquid-liquid extraction (LLE) using methyl tert-butyl ether and analyzed by LC-ESI-MS/MS. The limit of quantification was 0.2 ng/mL and the method was linear in the range 0.2-60 ng/mL. The intra-assay precisions ranged from 1.6 to 7.7%, while inter-assay precisions ranged from 2.1 to 5.1%. The intra-assay accuracies ranged from 97.5 to 108.2%, and the inter-assay accuracies ranged from 97.3 to 107.0%. The analytical method was applied to evaluate the relative bioavailability of two pharmaceutical formulations containing 8 mg of ondansetron each in 25 healthy volunteers using a randomized, two-period crossover design. The geometric mean and respective 90% confidence interval (CI) of ondansetron test/reference percent ratios were 90.15% (81.74-99.44%) for C(max) and 93.11% (83.01-104.43%) for AUC(₀-t). Based on the 90% confidence interval of the individual ratios (test formulation/reference formulation) for C(max) and AUC(₀-inf), it was concluded that the test formulation is bioequivalent to the reference one with respect to the rate and extent of absorption of ondansetron.
22
Claudia Müller, Patrick Schäfer, Mylène Störtzel et al. · Journal of Chromatography B · 2002 · 253 citations
Yunsheng Hsieh, Madhu Chintala, Jacqueline Agans et al. · Rapid Communications in Mass Spectrometry · 2001 · 135 citations