Pacing and Clinical Electrophysiology · 1998 · 24 citations · 9 references
MedicineCardiologySurgeryCardiothoracic SurgeryElectrophysiologyInterventional CardiologyCardiac PacingHeart FailureImplantable Cardioverter-defibrillator CarePacemaker CareCardiac PacemakersImplantable Cardioverter DefibrillatorsInappropriate Shock ReductionLead DislodgementPacemaker ComplicationsPacemaker MalfunctionSubmuscular Implant Placement
Twiddler's syndrome is a highly recognized yet rare complication of pacemaker and cardioverter defibrillator (i.c.d.) implantation. We present a case in which persistent generator rotation resulted in lead dislodgment and inappropriate shocks in an initial ICD and recurrent lead fracture in a second ICD system. This case is unusual in that even with extensive surgical precautions including use of a Dacron pouch, generator rotation could not be prevented. Submuscular implantation and use of a smaller generator may prevent Twiddler's syndrome.
9
A Stretch Fabric Pouch for Implanted Pacemakers
Victor Parsonnet · Archives of Surgery · 1972 · 40 citations
Medicine, Cardiology, Surgery +17