Novel Membranes Amperometric Peroxide Biosensor Based on a Tetrathiafulvalene-p-Tetracyanoquinodimethane Electrode

Ulrich Korell, Ursula E. Spichiger

Analytical Chemistry · 1994 · 30 citations · 18 references

Concepts

Abstract

Amperometric biosensors for the detection of hydrogen peroxide are prepared by adsorbing peroxidase (POD, EC 1.11.1.7, lipophilized with caprylic aldehyde) to TTF-TCNQ/silicone oil paste electrodes. This is the first time a reductase is coupled to an organic conducting salt electrode. At -50 mV vs Ag/AgCl and pH 6.0, the current vs concentration function can be described by the enzyme kinetic Michaelis-Menten formalism. Stable signals are obtained within 10 s. The detection limit is typically in the low nanomolar range for H2O2. The enzyme stability under storage, standby, and various operation conditions is discussed.