Physical Review · 1925 · 16 citations · 4 references
EngineeringChemistryAnalytical ChemistryInstrumentationMetrologyAnalytical CalibrationCalibrationPressure MeasurementPressure CalibrationThermal CalibrationVapor Pressure
Mercury vapor pressure, 20\ifmmode^\circ\else\textdegree\fi{} to -80\ifmmode^\circ\else\textdegree\fi{}C.---A modified Buckley ionization gage is described with a range from ${10}^{\ensuremath{-}2}$ to 4\ifmmode\times\else\texttimes\fi{}${10}^{\ensuremath{-}9}$ mm. This was used to measure the vapor pressure of mercury. Correction for thermal effusion was made, ranging from 25 percent at -80\ifmmode^\circ\else\textdegree\fi{}C to zero at 0\ifmmode^\circ\else\textdegree\fi{}C. The gage was calibrated at 0\ifmmode^\circ\else\textdegree\fi{}C from Knudsen's value. The results correspond to the equations ${log}_{10}P(\mathrm{liquid})=8.86\ensuremath{-}3.44(\frac{{10}^{3}}{T})$; ${log}_{10}p(\mathrm{solid})=9.35\ensuremath{-}3.55(\frac{{10}^{3}}{T})$.Heat of vaporization of mercury, 20\ifmmode^\circ\else\textdegree\fi{} to -80\ifmmode^\circ\else\textdegree\fi{}.---From a large-scale plot of $logp$ AGAINST $\frac{1}{T}$ the heats of vaporization were calculated from the relation $\ensuremath{\lambda}=\ensuremath{-}\frac{Rd(logp)}{d(\frac{1}{T})}$. Above the melting point the results correspond to the equation $\ensuremath{\lambda}=15,900\ensuremath{-}10t$ where $t$ is temperature centigrade. A slight discontinuity in slope of the ($logp,\frac{1}{T}$) curve gives at the melting point a sudden increase in the heat of vaporization of some 600 calories, which is in good agreement with the heat of fusion by direct experiment.
4