The measured radiation dose to the phantom was identical for all pitch selections on the multislice helical CT system we tested. This unexpected result was because of an automatic proportionate increase in the tube current when the pitch selection was increased. Radiologists and physicists should exercise caution when extrapolating dose reduction strategies from single-slice to multislice helical CT systems, and they must acquire a detailed understanding of the multislice helical CT scanner of their chosen manufacturer.