Magnetic Resonance in Medicine · 1985 · 77 citations · 5 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage RestorationMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMedical ImagingArtifact ReductionArtifacts In Magnetic Resonance ImagingFourier ReconstructionImage ArtifactsMagnetic ResonanceTruncation
A method has been developed to reduce a magnetic resonance image artifact arising from the Fourier transformation of data truncated in the phase-encoding direction. The discontinuity is eliminated between the phase of the complex data sample associated with the maximum negative phase-encoding gradient and the phase of the maximum positive gradient sample. Images retain sharper edges than if low-pass filtering is employed. The method is demonstrated on clinical images, then an explanation of the method is presented.
5
The Fourier Transform and its Applications
K.W. Cattermole · Electronics and Power · 1965 · 1.4K citations