Publication | Closed Access
Major stent deformation / pseudofracture of 7 Crown Endeavor / Micro Driver stent platform: incidence and causative factors
39
Citations
9
References
2011
Year
Stent deformation/pseudofracture of the 7 crown Endeavor/Micro Driver platform occurred in 1.8% of cases that were post-dilated. Of these, 36% experience MACE. Deformation can be reproduced on the benchtop. Operators should be aware of the complication, and develop strategies to deal with it.
| Year | Citations | |
|---|---|---|
Page 1
Page 1