Applied Optics · 1999 · 62 citations · 16 references
Natural SciencesBiomedical EngineeringBiophysicsBiomedical OpticsBiophotonicsOpticsLight MicroscopyMultiphoton MicroscopyOptical ImagingRefractive IndexNumerical AperturePoint Spread FunctionRefractive Index MatchingRefractive Index MismatchTwo-photon Excitation MicroscopyTwo-photon Fluorescence ImagingTwo-photon Fluorescence MicroscopyTwo-photon Microscopy
The detrimental effects of a refractive-index mismatch on the image formation in a two-photon microscope were investigated. Point-spread functions (PSF's) were recorded with an oil-immersion objective numerical aperture (NA) of 1.3 and a water-immersion objective NA of 1.2 in an aqueous sample at different depths. For the oil-immersion objective the enlargement of the PSF volume with increasing depth yields an axial and a lateral loss in resolution of approximately 380% and 160%, respectively, at a 90-microm depth in the sample. For the water-immersion objective no resolution decrease was found. Measurements on a thick aqueous biofilm sample shows the importance of matching the refractive index between immersion fluid and sample. With a good match, no loss in image resolution is observed.
16
Two-Photon Laser Scanning Fluorescence Microscopy
Winfried Denk, James H. Strickler, Watt W. Webb · Science · 1990 · 9.4K citations